AM1760/1762 Secondary SPRT’s

AM1760 series Secondary Standard PRT provides our customers an affordable SPRT alternative for precision temperature measurement and calibration in labs and fields. This SPRT features accuracy of ±0.006°C at 0.01°C , short term stability of ±0.002°C and very low drift rate of less than 19 mK after 500 hours. Two different lengths of SPRTs are available at 12-inch and 20-inch. The sensing element is designed to protect the platinum sensing wire from contamination at high temperatures, giving the device a high level of stability and repeatability in performance. A uniquely designed support structure and filling material provides excellent balance between the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe fully meets ITS-90 criteria for reference thermometers.

  • Temperature range: -200 °C to 670 °C
  • Accuracy: ±0.006°C at 0.01°C
  • Long term drift: ±0.019°C
  • Short term stability: 0.002 °C
  • Temperature Coefficient 0.003925
  • W(Ga)>=1.11807
  • Customized dimensions available
SPECIFICATIONS
Temperature Range -200°C to 670°C
Resistance at 0 °C 1760 – Nominal 100 Ω 1762 – Nominal 25 Ω
Temperature Coefficient 0.003925 Ω/ Ω/°C
Accuracy ±0.007°C at -196 °C ±0.006°C at 0.01 °C ±0.015°C at 420 °C ±0.025°C at 660 °C
Drift* ±0.004°C at 0 °C after 100 hours at 660 °C
Short Term Stability ±0.002°C
Thermal Shock ±0.002°C after 10 times thermal cycles from minimum to maximum temperatures
Self-heating 0.0015 °C at 1 mA current
Response Time 9 seconds for 63{f2d75a6b7d0fe049c173b1b8e8fb128a0cddb2c9179e3e439d4968f3513d2996} response to step change in water moving at 3 feet per second
Measurement Current 0.5 mA or 1 mA
Sensor Length 42 mm
Sensor Location 5 mm from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Inconeltm
Dimension 1760/1762-12: 0.25 inch X 12 inch (6.35 mm X 305 mm) 1760/1762-20: 0.25inch X 20 inch (6.35 mm X 500 mm)
External Leads Teflontm –insulated copper wire, 4 leads, 2.5 meters
Handle Dimension 15mm (OD) X 65 mm (L)
Handle Temperature Range** -50°C to 180°C
Calibration Options 1760/1762-12-T, Secondary SPRT with NIST traceable calibration and data 1760/1762-20-T, Secondary SPRT with NIST traceable calibration and data
Model Description
9001 Wooden Carrying Case for 1760/1762-12
9002 Wooden Carrying Case for 1760/1762-20

AM1760-1762 brochure (PDF Download 470Kb)